Qualified by Electricite de France to meet the requirements of RCC-E K3 applications, the Rosemount 3153K Nuclear Qualified Pressure Transmitter is an analog transmitter designed for operating nuclear reactors in France. Featuring steam pressure/temperature performance, this device is qualified per RCC-E 2002 and withstands radiation levels of 20 kGy (2 Mrad) TID gamma radiation and vibration levels up to 50 m/s2 at 16 Hz.
Output Signal: 4-20mA
Rangedown: Up To 10:1 (5:1 for range 1)
Measurement Range: Up to13.79 MPa (2000 psi) Differential, Up to 27.58 MPa (4000 psig) Gauge, Up to 27.58 MPa (4000 psig) Absolute
Housing Material: Aluminum
Process Wetted Material: 316 SST
Seismic Rating: Vibration levels up to 50 m/s2 at 16 Hz
Radiation Rating: 20 kGy (2 Mrad) TID gamma radiation
Steam/Pressure Temperature: HELB 115° C (239° F) peak temperature
Qualification Standard: RCC-E 2002
Product Literature: The Product Data Sheets and Manuals posted on this web-site are intended for general product information only, with errata included. They may not represent the latest revision or contain information relating to recent product notifications, if any.
Contact Information: Please contact Rosemount Nuclear Instruments, Inc. at 1-952-949-5200 or[email protected]for the most current information available