Insulation Resistance: 1000 MΩ minimum insulation resistance when measured at 500VDC at room temperature
Certifications/Approvals: Hazardous location, see full specs for complete list of certifications
Lead Wires: 24 AWG wire, FEP insulated. Color coded per IEC 60584 or ASTM E230.
Insulation Resistance: 1000 MΩ minimum insulation resistance when measured at 500VDC at room temperature
Certifications/Approvals: Hazardous location, see full specs for complete list of certifications
Input: Eight independently configurable inputs, including 2- and 3- wire RTDs, thermocouples, mV, 2- and 3- wire ohm, and 4-20 mA signals
Output Signal: FOUNDATION™ Fieldbus protocol
Housing: Aluminum, or stainless steel junction box
Diagnostics: Measurement validation
Certifications/Approvals: Hazardous location, marine type, see full specs for complete list of certifications
Thermowell Calculations: Wake frequency calculation (WFC) performed by certified engineers, order R21 for official report
Design Modifiers: Root, bore and tip diameters can be customized for any application
Certifications/Approvals: Material certification, NACE, CRN (see full specs for complete list of certifications)
Product Tests: External pressure test, internal pressure test, dye penetration, X-ray, PMI (see full specs for complete list)